ID30A-3 (MASSIF-3) – a beamline for macromolecular crystallography at the ESRF with a small intense beam

Antoine Royant
Olof Svensson
  • Fonction : Auteur
  • PersonId : 1022003

Résumé

ID30A-3 (or MASSIF-3) is a mini-focus (beam size 18 µm × 14 µm) highly intense (2.0 × 1013 photons s-1), fixed-energy (12.81 keV) beamline for macromolecular crystallography (MX) experiments at the European Synchrotron Radiation Facility (ESRF). MASSIF-3 is one of two fixed-energy beamlines sited on the first branch of the canted undulator setup on the ESRF ID30 port and is equipped with a MD2 micro-diffractometer, a Flex HCD sample changer, and an Eiger X 4M fast hybrid photon-counting detector. MASSIF-3 is recommended for collecting diffraction data from single small crystals (≤15 µm in one dimension) or for experiments using serial methods. The end-station has been in full user operation since December 2014, and here its current characteristics and capabilities are described.
Fichier principal
Vignette du fichier
ig5090.pdf (1.29 Mo) Télécharger le fichier
Origine Publication financée par une institution

Dates et versions

hal-02864606 , version 1 (26-11-2020)

Identifiants

Citer

David von Stetten, Philippe Carpentier, David Flot, Antonia Beteva, Hugo Caserotto, et al.. ID30A-3 (MASSIF-3) – a beamline for macromolecular crystallography at the ESRF with a small intense beam. Journal of Synchrotron Radiation, 2020, 27 (3), pp.844-851. ⟨10.1107/S1600577520004002⟩. ⟨hal-02864606⟩
207 Consultations
212 Téléchargements

Altmetric

Partager

More